64
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Installation of the Operating Unit
6.3.5
Switching outputs (threshold value alarm, status)
Electronic properties
●
The switching outputs are optical coupler outputs (
●
The logics required by the user is set at the manufacturer's works.
Functions
NOTICE:
Only connect each switching output to 24 VDC/100 mA max.
Only connect inductive loads if suppressor diodes are integrated or
installed for protection against induced voltages.
Pay attention to the switching logic (
Table 4
Switching outputs: Functions
Name
Function
Measured value valid
Activated if the current measured value of the EuroFID is valid.
Calibration
Activated as long as a calibration is running.
1
1
Corresponds to the status signal “Check” of the NAMUR recommendations
Malfunction
Deactivated if there is an internal fault in the function of the EuroFID.
2
2
Display of the possible cause
Service requirement
Activated if the EuroFID can still perform measurement, but several internal
conditions are no longer in the operating range for valid measuring values.
3
4
3
Corresponds to the status signal “Service requirement” of the NAMUR recommendations
4
A corresponding message is shown on the display.
Alarm if the threshold
value is not reached
(L2)
Activated as long as the measured value is outside the set threshold value.
Option:
L2 and H2 remain activated until they are reset by means of a menu
P. 124, §11.11.9; reset
Instead of L2/L1/H1/H2 two pre- one alarm can be configured to
indicate when a threshold value is exceeded (H0/H1/H2
Pre-alarm if the
threshold value is not
reached (L1)
Pe-alarm if the
threshold value is
exceeded (H1)
Alarm if the threshold
value is exceeded (H2)
Table 5
Switching outputs: switching logic
Name
deactivated status
activated status
Measured value valid
Measurement mode is faulty
Measuring mode is running
Calibration
No calibration
Calibration is running
Malfunction
Malfunction!
1
1
With this switching logic, “Malfunction” is displayed if the mains supply fails.
–
Service requirement
No service requirement
Service requirement!
Alarms/pre-alarms
No alarm
The current measured value is outside
the respective threshold value
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...