Clearing Malfunctions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
179
38
A n - B W t o l A e i 2 "
S E R V I C E R E Q U I R E M E N T
A n a l o g i n p u t - 2 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
The input signal of
analog input 2 is faulty.
The measured values
can be corrupted –
depending on the use
of the analog input
Disappears as soon
as the cause is
eliminated.
Check the input signal
source (status, current
signal).
Check the electrical
connection.
39
A n - R l f a i A i n 2 "
S Y S T E M E R R O R
A n a l o g i n p u t - 2 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
40
A n - R l l i m A i n 2 "
S Y S T E M E R R O R
A n a l o g i n p u t - 2 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
41
A n a l . R e s t a r t "
M e s s a g e
T h e m i c r o p r o c e s s o r
o f t h e a n a l y z e r u n i t
h a s s t a r t e d .
Temporary malfunction
in the microprocessor
system of the analyzer
unit (message only
appears in the error
archive).
–
If this malfunction occurs
frequently:
Notify the after-
sales service of the
manufacturer.
42
A n - B W t o l A e i 3 "
S E R V I C E R E Q U I R E M E N T
A n a l o g i n p u t - 3 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
The input signal of
analog input 3 is faulty.
The measured values
can be corrupted –
depending on the use
of the analog input
Disappears as soon
as the cause is
eliminated.
Check the input signal
source (status, current
signal).
Check the electrical
connection.
43
" A n - R l f a i A i n 3 "
S Y S T E M E R R O R
A n a l o g i n p u t - 3 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
44
A n - R l l i m A i n 3 "
S Y S T E M E R R O R
A n a l o g i n p u t - 3 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
45
E P R O M A N A E R R "
S Y S T E M E R R O R
P e r m a n e n t
m e m o r y t e s t
a n a l y z e r .
E r r o r o c c u r r e d
d u r i n g E P R O M t e s t
r e a d .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
If this malfunction occurs
frequently:
Notify the after-
sales service of the
manufacturer.
47
A n - R l f a i A i n 4 "
S Y S T E M E R R O R
A n a l o g i n p u t - 4 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
The input signal of
analog input 4 is faulty.
The measured values
can be corrupted –
depending on the use
of the analog input
Disappears as soon
as the cause is
eliminated.
Check the input signal
source (status, current
signal).
Check the electrical
connection.
48
A n - R l l i m A i n 4 "
S Y S T E M E R R O R
A n a l o g i n p u t - 4 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
49
R A M _ A N A _ E R R .
S Y S T E M E R R O R
P e r m a n e n t
m e m o r y t e s t
a n a l y z e r .
E r r o r o c c u r r e d
d u r i n g R A M t e s t
r e a d / w r i t e .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
If this malfunction occurs
frequently:
Notify the after-
sales service of the
manufacturer.
N Spontaneous message
Message text in the error archive
Significance
Duration of the
message
Countermeasures
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...