“Maintenance” Sub-menu Functions
11.11.3
Setup filter (averaging for the “Measured value” analog output)
Function
It is possible that a plotting device is connected to the “Measured value” analog output
which does not record the measured value continuously, but only at regular intervals – e.g.
a printing recorder. It can be advantageous to average the analog measurement signal over
the respective time interval; otherwise the plotter would register momentary values which
may not be representative.
To achieve this effect, the “Setup filter” can be programmed.The setting value is the time
constant of sliding averaging (time-related integration).
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
I/O-
Parameters
Setup filter
2
Press
or
.
➝
Entry field is highlighted.
3
Enter the required value.
4
Press
or
.
➝
E n d
is highlighted.
5
Press .
●
The “Setup filter” setting only affects the “Measured value” analog output
(
).
●
Standard value for emission measurement: 30 s.
There is also a similar setting option
– for the measurement display
– for the “Reference value” analog output
S y s t e m P a r a m e t e r s
S e t u p f i l t e r
I n t e g r a t i o n t i m e
3 0 . 0 s
u p : 0 . 0 s
u p t o : 6 0 0 . 0 s
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Menu title
Entry field
Minimum value
Maximum value
Current measured value
Current measuring range (measuring range end value)
E n d
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...