EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Installation of the Operating Unit
6.3.3
Analog outputs
Functional principle
●
The analog outputs supply a load-independent current signal.
●
The electronic measuring range is 0 … 20 mA or 4 … 20 mA (setting
).
●
The electronic measuring range corresponds to the current measuring range or a
permanently set range (setting
Malfunction display
If the current measured values are not valid, the following measurement signal is output:
Analog output “Measured value”
The current signal of the “Measured value” analog output corresponds to the current
measured value in the currently selected measuring range.
●
The current measuring range can be signaled by the “Reference value” analog output.
●
Sliding averaging can be activated (
●
The function is variable during calibration (
).
Analog output “Reference value”
The function of the “Reference value” analog output depends on the set measuring range
mode (
CAUTION:
Possible danger for connected devices/stations
When the operating unit is switched on, the analog outputs emit a current of
22 mA for a short time (= 110 % of the measuring range).
Before switching the operating unit off:
If necessary for safety reasons,
inform connected stations and/or deactivate connected devices (e.g. alarm
indicator).
Electronic measuring range
Measurement signal in the event of invalid
measured values
1
1
Status output “Measured value valid”
0 … 20 mA
0 mA
4 … 20 mA
2 mA
Output signal of the “Reference value” analog output
Range select mode = “autoAOu”
Other measuring range modes
Signal for marking of the measuring
range (
–
In measuring mode:
Measurement signal
same as the “Measured value” analog output
1
–
During calibration:
The last measured value
which was measured before start of calibration
1
Activation of sliding averaging
Table 3
Measuring range signal on the “Reference value” analog output
Significance
Output signal in mA
Electron. measuring range=
0 … 20 mA
Electron. measuring range=
4 … 20 mA
Measuring range 1
2
5.6
Measuring range 2
4
7.2
Measuring range 3
6
8.8
Measuring range 4
8
10.4
Measuring range 5
10
12.0
Measuring range 6
12
13.6
Measured value is invalid
<1
<4.8
> 13
> 14.4
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...