“Maintenance” Sub-menu Functions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
123
11.11.6
Output range
Function
The following properties of the “Measured value” analog output (
) can be
modified:
●
Electronic measuring range
●
Reference of the electronic measuring span (measuring range or fixed setting)
●
Function during calibration
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
I/O-
Parameters
Output range
2
Press
or
, until the respective field is highlighted.
3
With selection boxes: Press
until the required setting is displayed.
With entry fields:
Enter the required value.
To finish:
4
Press
or
, until
E n d
is highlighted.
5
Press .
Table 12
Settings for the “Measured value” analog output
C u r r e n t
0 - 2 0 m A
Electronic measuring span = 0 … 20 mA
4 - 2 0 m A
Electronic measuring span = 4 … 20 mA
1
1
The output signal 2 mA means malfunction (for a detailed explanation refer to
M o d e
a s M R
The electronic measuring span corresponds to
the current measuring range (0 … 100 %)
R a n g e
The electronic measuring span corresponds to
the set values (
f r o m
…
t o
…)
f r o m
2
2
Adjustable if “Range” is selected as the mode
Beginning value of the bar scale
t o
End value of the bar scale
Z e r o G a s A n . O u t p
Y
During calibration the zero gas value is output.
N
During calibration the measured value “0” is
output during calibration.
S y s t e m P a r a m e t e r s
O u t p u t r a n g e
O u t p u t r a n g e
C u r r e n t 4 - 2 0 m A
M o d e = r a n g e
u p 0
u p t o 1 0 0 0
p p m
Z e r o G a s - A n . O u t p N
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Menu title
Selection box 1
Selection box 2
Entry field (only with the “Range” mode)
Entry field (only with the “Range” mode)
Selection box 3
Current measured value
Current measuring range (measuring range end value)
Signal range of the “Reference value” analog output
E n d
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...