“Maintenance” Sub-menu Functions
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
121
11.11.4
Median value time (averaging for the “Reference value” analog output)
Function
The “Median value” function performs the same function for the “Reference value” analog
output as the “Setup filter” function for the “Measured value” analog output (
).
Procedure
1
Open the menu: Main menu
Maintenance
System Parameters
User level
I/O-
Parameters
Median value time
2
Press
or
.
➝
Entry field is highlighted.
3
Enter the required value.
4
Press
or
.
➝
E n d
is highlighted.
5
Press .
S y s t e m P a r a m e t e r s
M e d i a n v a l u e t i m e
A v e r a g e t i m e
1 8 0 0 . 0 s
u p : 0 . 0 s
u p t o : 1 8 0 0 . 0 s
C h a n g e o r
6 1 . 6 m g C
R a n g e . . . 1 0 0
Menu title
Menu title
Entry field
Minimum value
Maximum value
Current measured value
Current measuring range (measuring range end value)
E n d
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...