178
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
Clearing Malfunctions
25
C l o c k e r r o r
S Y S T E M E R R O R
T h e c l o c k i s
d e f e c t i v e a n d m u s t
b e c h a n g e d .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
EuroFID Switch off and on
once more.
If unsuccessful:
Have the
EuroFID serviced by the
after-sales service of the
manufacturer.
26
S e r v o d a t e n f e
S Y S T E M E R R O R
D u r i n g p r e s s u r e
c o n t r o l a n e r r o r h a s
o c c u r r e d . P l e a s e
c h e c k p r e s s u r e v a l u e
a n d d a t a l i n e t o t h e
a n a l y z e r a s s e m b l y .
Internal malfunction.
The measured values
are not valid. The
individual settings on
the device are not
effective.
Can only occur
during
manufacturing.
Disappears as soon
as the cause is
eliminated.
Have the EuroFID serviced
by the manufacturer's
after-sales service.
27
Q u e u F u l l S y s t
S E R V I C E R E Q U I R E M E N T
P r i n t e r s p o o l e r i s
f i l l e d . P l e a s e c h e c k
t h e p r i n t e r .
Data output is faulty.
Measurement is not
interrupted.
Disappears as soon
as the cause is
eliminated.
Check the status of the
connected printer.
Check the connection to
the printer (plug
connections, cables).
28
Q u e u F u l l M v a l
S E R V I C E R E Q U I R E M E N T
T h e m e a s u r e d v a l u e
b u f f e r i s
f i l l e d . P l e a s e c h e c k
t h e p r i n t e r .
Data output is faulty.
Measurement is not
interrupted.
Disappears as soon
as the cause is
eliminated.
Check the connected
systems.
Check the electrical
connections (plug
connections, cables).
29
Q u e u F u l l L i s t
S E R V I C E R E Q U I R E M E N T
T h e l i s t b u f f e r i s
f i l l e d . P l e a s e c h e c k
t h e p r i n t e r .
30
S p o o F u l l
S E R V I C E R E Q U I R E M E N T
P r i n t e r s p o o l e r i s
f i l l e d . P l e a s e c h e c k
t h e p r i n t e r .
Data output is faulty.
Measurement is not
interrupted.
Disappears as soon
as the cause is
eliminated.
Check the status of the
connected printer.
Check the connection to
the printer (plug
connections, cables).
31
P r n t O u t _
S E R V I C E R E Q U I R E M E N T
T h e p r i n t e r i s n o t
r e a d y . P l e a s e c h e c k
t h e p r i n t e r .
32
P r n t O u t _ P a p r
S E R V I C E R E Q U I R E M E N T
T h e p r i n t e r h a s n o
p a p e r . P l e a s e f i l l
u p p a p e r .
33
C l o c k r a m e r r
S E R V I C E R E Q U I R E M E N T
T h e p a r a m e t e r s i n
t h e c l o c k c h i p a r e
l o s t . P l e a s e c h e c k
t h e c l o c k c h i p .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
Reset the time of the
internal clock (
).
Then calibrate the device
34
A n - B W t o l A e i 1
S E R V I C E R E Q U I R E M E N T
A n a l o g i n p u t - 1 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
The input signal of
analog input 1 is
faulty.The measured
values can be
corrupted – depending
on the use of the
analog input (
).
Disappears as soon
as the cause is
eliminated.
Check the input signal
source (status, current
signal).
Check the electrical
connection.
35
A n - R l f a i A i n 1 "
S Y S T E M E R R O R
A n a l o g i n p u t - 1 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
The input signal of
analog input 1 is faulty.
The measured values
can be corrupted –
depending on the use
of the analog input
(
).
Disappears as soon
as the cause is
eliminated.
Check the input signal
source (status, current
signal).
Check the electrical
connection.
36
A n - R l l i m A i n 1 "
S Y S T E M E R R O R
A n a l o g i n p u t - 1 -
A / D - C o n v e r t e r i n p u t
i s o u t o f t o l e r a n c e .
P l e a s e c h e c k t h e
c o n n e c t e d u n i t .
37
E P R O M _ E R R O R "
S Y S T E M E R R O R
P e r m a n e n t
m e m o r y t e s t .
E r r o r o c c u r r e d
d u r i n g E P R O M t e s t
r e a d .
Internal malfunction.
Measurement is
interrupted.
Disappears as soon
as the cause is
eliminated.
Have the EuroFID serviced
by the manufacturer's
after-sales service.
N Spontaneous message
Message text in the error archive
Significance
Duration of the
message
Countermeasures
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...