Installation of the Analyzer Unit
EuroFID · Operating Instructions · 8011094 V2.3 · © SICK AG
41
4.2.2
Gas connections
Install the gas connections in the same manner as described for the analyzer type “wall
housing” (
4.2.3
Electrical connections
Connect the joint plug connector of both connection cables to the terminal box
(integrated into the operating unit).
On devices with the “flow barrier” option:
Connect the solenoid valve to the terminal
box (
Fig. 6
Analyzer unit “19" slide-in enclosure” type: Connections
Auspuff
Outlet
3 bar
H2
/He
Fuelgas
3 bar
Kalibriergas
Calibr.gas
3 bar
Instr.Luft
Instr. air
System
Printer
Drucker
RS 232 C
Datenleitung
Data cable in
Analyzer Assembly
Terminalbox
AC 115V/230V
Warning
Achtung
Air filter must be cleaned in regular intervals.
Luftfilter muss regelmäßig gereinigt werden.
Caution
Achtung
Power
Netz
Fuses
Sicherungen
4A MT
Analog and Digital I/O
Analog
e und digitlale
Ein-/A
usgabe
Electric shock hazard. Do not remove cover.
Servicing only by qualified personnel.
Nicht unter Spannung öffnen.
Wartung nur durch qualifiziertes
Personal.
3 bar
Kalibriergas
Calibr.gas
3 bar H
2
He
Brenngas
Fuelgas
Auspuff
Outlet
4 bar
Instr.Luft
Instr. air
Bypass
Outlet
Sample gas
*
*
*
* ±0.2 bar
Summary of Contents for EuroFID
Page 10: ...Contents 10 EuroFID Operating Instructions 8011094 V2 3 SICK AG ...
Page 18: ...18 EuroFID Operating Instructions 8011094 V2 3 SICK AG Important Information ...
Page 80: ...80 EuroFID Operating Instructions 8011094 V2 3 SICK AG Start up ...
Page 88: ...88 EuroFID Operating Instructions 8011094 V2 3 SICK AG Operation General ...
Page 100: ...100 EuroFID Operating Instructions 8011094 V2 3 SICK AG Main Menu Functions ...
Page 144: ...144 EuroFID Operating Instructions 8011094 V2 3 SICK AG Maintenance Sub menu Functions ...
Page 154: ...154 EuroFID Operating Instructions 8011094 V2 3 SICK AG Calibration ...
Page 158: ...158 EuroFID Operating Instructions 8011094 V2 3 SICK AG Shutdown ...
Page 182: ...182 EuroFID Operating Instructions 8011094 V2 3 SICK AG Clearing Malfunctions ...