Chapter
6
Operating Instructions
CLV 490 Bar Code Scanner
6-46
©
SICK AG · Division Auto Ident · Germany · All rights reserved
8 009 993/O824/20-12-2004
Operation
4.
Choose C
LOSE
to close the window.
If the parameters in the CLV were modified as a result of the teach-in procedure, CLV-
Setup asks you whether you want to copy and display the modified parameter set by
uploading it from the CLV.
5.
Confirm the query with Y
ES
.
The dialog box closes automatically.
The CLV resumes Reading mode and the "Device Ready" LED lights up.
Displaing the taught-in background:
The new background distance profile can be uploaded and displayed in CLV-Setup via the
terminal interface of the CLV. CLV-Setup displays the background without any parameterized
limitations ("Autofocus space").
1.
Choose the R
EADING
C
ONFIGURATION
tab.
2.
In the A
UTOFOCUS
section, click the S
HOW
B
ACKGROUND
button.
The S
HOW
B
ACKGROUND
dialog box appears.
An example of a distance profile is shown in
Note
The CLV adjusts the entire profile of the taught-in background by minus 100 mm (3.94 in),
that is, the entire background curve is moved toward the CLV by this amount (also in the
display in CLV-Setup) to compensate measurement inaccuracies. The parameterized limits
of the applicable autofocus space are not corrected. The CLV also shifts the edges of the
measured profile toward the center by 2 angular elements.
If no background has been taught in, the CLV uses the default background (2,200 mm (
86.7 in) across the entire distance profile).
Fig. 6-12:
CLV-Setup: Display of th learned background