This unit has
self-diagnostic functions.
So
when abnormality
is detected, one
of
the following error messages is displays along
with the output of alarm sound.
mode
1.
System error (Failure of the main unit)
(1)
R O M
F A
I
L U R E
(2)
RAM error
(3)
Home position error
(4)
Leakage error
Overheat error
2.
Set value loss error
It is displayed when a ROM error is detected.
If
this error is displayed, turn the power
of
the unit
off
and contact
your Shimadzu Service Representative.
balance
range
mode
R A M
F A I
L U R E
It is displayed when a
RAM
error is detected.
If
this
is displayed,
turn
the power of the unit
off
and contact
your Shimadzu Service Representative.
balance
range
mode
E R R
Z G
H O M E P O S
It is displayed when the home position sensor
of
the motor for
driving the xerography does not properly work.
If this error is displayed,
turn
the power
of
the unit
off
and contact
your Shimadzu Service Representative.
balance
range
mode
E R R L E A K
6 5 5 3 5
It
is
displayed when the leakage occurs.
Remove the cause
of
leakage, and wipe
off
the liquid on the leak
sensor. When the displayed value comes to
or
less, press the
key to stop the alarm and use the leak sensor again.
balance
range
mode
E R R O V E R H E A T
It is displayed when the temperature
of
the measuring point rises to
63°C
and
over.
If this
error
is displayed,
turn
off
the power
of
the
unit and contact your Shimadzu Service Representative.
balance
range
mode
N O T
P R O T E C T E D
It is displayed
if
the parameters and time program set previously
are lost at powering up.
Press
key and the unit becomes
operational. Set parameters
again.
10-4
RID-1OA
Summary of Contents for RID-10A
Page 19: ...Chapter 2 Parts List c v 3 2 1 Parts and Accessories 2 2 1CONTENTS 1 2 1 RID 1OA...
Page 63: ...c Chapter 6 Initial Performance Test 1 6 1 System Performance Check 6 2 6 1 RID 1OA...
Page 102: ...Chapter I 1 Specifications 11 1 RID 10A Specifications 11 2 I CONTENTS I 11 1 RID 1OA...
Page 105: ...Parts I 12 1 Consumablesand Repair Parts 12 2 12 2 Optional Parts 12 4 12 1 RID 1OA...
Page 109: ...Chapter 13 Reference I i L 13 1 Solvent Characteristics 13 2 I 13 1 RID 1OA...