9.1
Component Validation
4.
Checking the span
Content of inspection
This test checks whether the refractive index value
is
accurately
measured and displayed by testing the standard sample whose
index is already
known.
The test requires the following reagents and equipment.
Distilled water
Standard sample (Glycerin solution)
Syringe
Data processor
Refer to the section
8.2
Span Adjustment.
Test procedure
Set the parameters
of
the detector
as
follows.
CELL
TEMP:
polarity:
+
mode: A (When performing mode A span adjustment)
Set the detector
so
that the plot full scale of the data processor
becomes
128
x
AUX RANGE of the
2
(or
4)
of the Chromatopac:
7
(or
9)
Use the syringe and adapter to deliver distilled water to the
(2)
(3)
inlet port. During delivery,
press
the
and R
keys
to flow the water to the reference flow lines.
After the inside of the sample
cell is replaced
with distilled water,
OFF
by pressing
the
and
keys again.
When the baseline is stabilized, press the
and
keys to perform optical balance adjustment.
Press the
key and record the baseline level when the
cell is filled with distilled water.
Inject the standard solution from
the
inlet port. The baseline
moves up ward by
80% of
the full scale.
(4)
( 5 )
(6)
(7)
Record the baseline level when the cell is filled with
standard solution. When
the
Chromatopac is used, press the
PLOT
key to record the baseline.
(8)
Read the difference of the baseline level between the
distilled water and the standard solution, and convert the
value into
unit.
9-4
Summary of Contents for RID-10A
Page 19: ...Chapter 2 Parts List c v 3 2 1 Parts and Accessories 2 2 1CONTENTS 1 2 1 RID 1OA...
Page 63: ...c Chapter 6 Initial Performance Test 1 6 1 System Performance Check 6 2 6 1 RID 1OA...
Page 102: ...Chapter I 1 Specifications 11 1 RID 10A Specifications 11 2 I CONTENTS I 11 1 RID 1OA...
Page 105: ...Parts I 12 1 Consumablesand Repair Parts 12 2 12 2 Optional Parts 12 4 12 1 RID 1OA...
Page 109: ...Chapter 13 Reference I i L 13 1 Solvent Characteristics 13 2 I 13 1 RID 1OA...