9.1
Component Validation
(3)
Select the
[6.
MONITOR] on the main menu
of
the
to open the monitor screen.
(Example
of MONITOR
screen]
40.0
7566
BALANCE
10
86
(4)
Record the value
of
as
the actual temperature.
Acceptance criteria
ACT TEMP value is within
+
-
0.1
of
CELL
TEMP.
The CELL
TEMP
value must be 12
higher than room temperature.
4.
Checking the span
Content
of
inspection
This test checks whether the refractive index value is accurately
measured and displayed by testing the standard sample whose
index is already
known.
The test requires the following reagents and equipment.
Distilled water
Standard sample (Glycerin solution)
Syringe
Data processor
Refer to the section 8.2 Span Adjustment.
Test procedure
(1)
Set the parameters
of
the detector
as
follows.
LOCAL:
(*)
CELL
40
(
indicated
as
OFF)
polarity:
+
mode: A (When performing mode A span adjustment)
*This test must be performed in LOCAL mode not in
REMOTE
mode. After changing the operation mode, other
parameters, such
as
CELL
TEMP,
polarity, and mode,
should be set directly to the detector.
Set the detector
so that the plot full scale of the data processor
becomes 128
x
AUX
RANGE
of the detector: 2 (or
4)
of the Chromatopac:
7
(or
9)
(2)
9-7
RID-1OA
Summary of Contents for RID-10A
Page 19: ...Chapter 2 Parts List c v 3 2 1 Parts and Accessories 2 2 1CONTENTS 1 2 1 RID 1OA...
Page 63: ...c Chapter 6 Initial Performance Test 1 6 1 System Performance Check 6 2 6 1 RID 1OA...
Page 102: ...Chapter I 1 Specifications 11 1 RID 10A Specifications 11 2 I CONTENTS I 11 1 RID 1OA...
Page 105: ...Parts I 12 1 Consumablesand Repair Parts 12 2 12 2 Optional Parts 12 4 12 1 RID 1OA...
Page 109: ...Chapter 13 Reference I i L 13 1 Solvent Characteristics 13 2 I 13 1 RID 1OA...