F-
polarity
mode
Span adjustment should be performed over each mode of A,
P
and
L.
When a Chromatopac is used, refer to
8.2.1,
and when
an
LC
workstation is used, refer to
Inject the standard solution whose refractive index is known, and
adjust the detector output
so
that the signal intensity can be
Following
is
the description for mode A span adjustment.
+
A
(When performing mode A span adjustment)
I
Items to be prepared
I
I
Chromatopac
or LC
workstation
.
Adapter
Water
(HPLC
grade)
Refractiveindex
*
Refractive index standard sample:
glycerin solution 0.872
Dissolve 43.6 g of glycerin
grade or equivalent) in the
of
water, and dilute this solution 50 times for the refractive index
standard solution. (Reference document
This
sample shows difference of
x
10"
from that of pure water.
8.2.1
When span adjustment
is
performed
using the Chromatopac
(1)
Turn
ON
the power.
(2)
Set the parameters of the detector
as
follows:
I
I
*
When the
OA
is
controlled from the system controller
SCL-
OA:
This
adjustment is performed by the detector independently. First, input
LOCAL
=
to
separate control
(link)
from the SCL. After that, input
each
set
value using the numerical keys of the detector main unit
8-3
RID-1OA
Summary of Contents for RID-10A
Page 19: ...Chapter 2 Parts List c v 3 2 1 Parts and Accessories 2 2 1CONTENTS 1 2 1 RID 1OA...
Page 63: ...c Chapter 6 Initial Performance Test 1 6 1 System Performance Check 6 2 6 1 RID 1OA...
Page 102: ...Chapter I 1 Specifications 11 1 RID 10A Specifications 11 2 I CONTENTS I 11 1 RID 1OA...
Page 105: ...Parts I 12 1 Consumablesand Repair Parts 12 2 12 2 Optional Parts 12 4 12 1 RID 1OA...
Page 109: ...Chapter 13 Reference I i L 13 1 Solvent Characteristics 13 2 I 13 1 RID 1OA...