Preventive measure 2.
Minimize the clearance of both inlet and outlet of vessel to prevent flame from entering the
vessel.
(1)
Cap with three holes for
18
liter and
4
liter cans
228-21
is available.
Connect clip to metallic
Minimize clearance
by
attaching a cap.
ground the metallic vessel.
Metallic
18
litter can
(A
plated can is recommended.)
Static electricity of the liquid
is released via vessel.
Fig.
3
Anti-Static Electricity Measures for Vessel
Preventive measure 3.
Do
not approach the vessel with charged objects including the human body.
Charging prevention measures for human body
a)
b)
Grounding
of human body
c) Make
floor conductive
Prevention of charging of shoes and clothes
Suitable products to be used for those measures a), b), and c) are available on the
market.
When persons who use no charge prevention measures approach dangerous sections,
they have to be grounded beforehand. (For example, they should contact grounded
metal by hand.)
Preventive measure
4.
Use pipes with inner diameter of
or more for waste liquid line for large flow rates.
Inclusion of bubbles in the
tube
may increase the amount of charging by ten times.
Check that there is no inclusion
of
air
via
tube
joints.
Summary of Contents for RID-10A
Page 19: ...Chapter 2 Parts List c v 3 2 1 Parts and Accessories 2 2 1CONTENTS 1 2 1 RID 1OA...
Page 63: ...c Chapter 6 Initial Performance Test 1 6 1 System Performance Check 6 2 6 1 RID 1OA...
Page 102: ...Chapter I 1 Specifications 11 1 RID 10A Specifications 11 2 I CONTENTS I 11 1 RID 1OA...
Page 105: ...Parts I 12 1 Consumablesand Repair Parts 12 2 12 2 Optional Parts 12 4 12 1 RID 1OA...
Page 109: ...Chapter 13 Reference I i L 13 1 Solvent Characteristics 13 2 I 13 1 RID 1OA...