background image

  

Mount the detector.

·

   

Wire  the  detector.

¹

 

  Test  the  detector.

¸

   

Configure  the  detector.

FG-1508 

Glassbreak 

Detector 

and 

Magnetic 

Contact 

Installation 

Instructions

Align the SWITCH with the

MAGNET to determine the

correct orientation for mounting.

The glassbreak detector and

magnetic contact housings must

be aligned and mounted

parallel to each other.

Configure jumpers W1-W4 to best suit the application.

Note:

 Some product versions do not have all of the

jumpers.

W1:

  Default: LEDs enabled.  To disable LED opera-

tion, remove the jumper from position W1 on the PCB.

W2:

  Default: alarm LED LATCH disabled. To enable

alarm LED LATCH, place a jumper across W2.

W3:

  Default: magnetic contact enabled. To disable

the magnetic contact, place a jumper across W3.

W4:

  Remove jumper for high sensitivity. Place jumper

across W4 for low sensitivity. Verify range by testing.

Note:

 Test the FG-1508 at least once each year.

Test the detector with the FG-701 Glassbreak

Simulator.

Activating Test Mode
1

Position the simulator within 5’ (1.5 m) of the

FG-1508 detector.

2

Switch the FG-701 simulator to ACTIVATE and

MANual modes.

3

Point the front of the simulator at the detector and

press the red start button.

The simulator buzzes, and the green LED on the

FG-1508 flashes once per second to indicate it is in

Test Mode.
If an FG-701 is not available, or if for any reason

remote activation cannot be used, use a small

screwdriver to short the test pads near location W4 on

the PCB. This activates Test Mode.

Note: 

 In Test Mode the LED disable jumper is

overridden.

Make sure to replace the front cover of the FG-1508

before testing.

Testing the FG-1508
1

Set the FG-701 switches to the TEST and FLEX

positions.

2

Press the red start button. The simulator “clicks”

on and starts an eight second armed period.

3

Position the FG-701 near the farthest point of the

glass and point it directly at the FG-1508. If

window coverings are present, close them fully

and hold the FG-701 behind them.

4

Generate a flex signal by carefully striking the

glass with a cushioned tool. The FG-701 responds

by producing a burst of glassbreak audio.

If both the flex and audio are received properly, the

red alarm LED on the FG-1508 lights.

IMPORTANT: 

 Some environmental factors may

reduce the detector activation range. If you do not see

the green LED flashing after pressing the red start

button, move closer to the detector and try again.

Testing the Audio Alone

You can also use the simulator in the MANual mode

to test audio alone. The blinking green LED on the

detector flickers when the simulator audio is received

correctly. (See the FG-701 Operating Instructions for

additional information.) Keep in mind that this is not

a complete test.

Exiting Test Mode

After testing, exit the Test Mode using the same

procedure for activating the Test Mode. The FG-1508

also automatically exits Test Mode five minutes after

the last glassbreak simulation.

Testing the Magnetic Contact (If Applicable)

With the magnetic contact enabled, the detector

signals an alarm to the control panel when the

normally closed magnetic contact opens. The LED

on the detector does not indicate the alarm condi-

tion. You must verify an open contact at the panel. To

reset the alarm relay, you must close the magnetic

contact (the door or window on which it is mounted

must be closed).

LEDs are enabled

Red alarm LED latches ON

when detector alarms

1,2

Magnetic contact is disabled

LEDs are disabled except

during power up test

Red alarm LED lights for 5

seconds  when  detector  alarms

Magnetic contact is

enabled

JUMPER

ON

OFF

W1

(LED)

W2

(LATCH)

W3

(CONTACT

DISABLE)

1

Latched alarm LED does not affect timing of alarm relay.

2

Reset the alarm LED by removing and restoring power, or by entering and

exiting test mode.

3

Always verify range of any sensitivity settings as described in the “Test” section.

Low  sensitivity

3

High sensitivity, 8’ max.

3

W4

(SENSITIVITY)

= Default Settings

FLEX

MAN

ACTIVATE

TEST

Red Start Button

FG-701 Simulator -- Top View

C

NC

V-

V+

TMPR

TMPR

Breakout tab

(both ends)

PCB wire entry

Terminal Block

W4 (Sensitivity)

W2 (Latch)

W3 (Contact Disable)

W1 (LED)

Test Pads

Route the wire through the wire entry hole in the PCB

housing and reconnect the housing to the mounting plate.

Cut and strip wire ends 1/4” (6.5 mm).

Wire the detector using 18-22 AWG (1.02mm-0.64mm).

Reverse polarity connections will not damage the detector.

Replace front cover.

For mounting guidelines, refer to the Supplemental Information on the back page.

Insert a screwdriver into the

slot at the end of the housing

and gently depress the front

cover latch. Lift and remove

the front cover.

Separate the mounting plate

from the glassbreak housing.

Insert a screwdriver into the

slot at the top of the PCB,

slightly twist, and gently push

downward on the latch.

To remove the front cover of

the magnetic contact housing,

insert a screwdriver into the

slot at the end of the back

cover and gently depress the

front cover latch. Lift and

remove the front cover.

Route the wires through the wiring channel.

Remove breakout tab and use an industry standard

4” (10 cm) tie wrap if a strain relief is required.

Secure the mounting plate and magnetic contact

housing with #6 (3.5 mm) screws (not provided).

To prepare for wiring, remove the appropriate

breakout tab in the PCB housing.

Magnet

Switch

TIP:  

Temporarily mount the detector in the intended

location and power it with a 9 V battery until testing

establishes effective range coverage.  If the 9 V

battery is low, it cannot supply sufficient power, and

the detector will not operate.

Reviews: