RID-1 OA
Specifications
Refractive index range
Measurement range
Linearity
(RIU)
1
Measuring
method
I
Deflection
1-1
0.01-500 x
(Analysis mode)
1-5000
x
and Large-scale Preparative modes)
5 x
(Analysis mode)
5
x
and
Large-scale Preparative modes)
Withstand pressure
Maximum operating flow rate
20 (with option installed
:
150)
analysis (A) mode)
Detector
:
0.4
Baseline noise
1
Detector internal capacity
Flow line switchover
Materials in contact
with
solvent
0.25
x
or less (filled with water, time constant 1.5
room temperature
analysis (A) mode)
Cell
:
1.9
IN
-
cell inlet
Cell
Cell outlet
-
OUT
:
280.2
Solenoid valve (pinch valve
type)
Maximum Operating pressure
:
quartz, fluoro-rubber, liquid
crystal
polymer
:
1
x
below (filled with water, time constant 1.5
room temperature
Cell
temperature setting
Zero
adjustment
30
-
60°C (by
OFF
Optical balance (optical
zero)
Auto-zero, baseline shifting Fine-zero)
Weight
Size W
x
Polarity switching
Possible
Time
Output (recorder output)
output)
Independent control of detector and control from system controller are both
possible.
Available parameters
:
range, marker, response, event, polarity, loop
RECORDER terminal,
RECORDER terminal
Selectable in
of 1
x
1
x
1
x
2.5
x
Safety measure
Bimetalic thermal breaker
I
Solvent leak sensor
Operating temperature range
4 - 3 5
Power
is classified
as follows according to supply voltage
20V
1
The
specifications
are determined according
to the
method below.
E1303-89
Practice for Refractive Index Detectors
Used
in LC.
11-2
Summary of Contents for RID-10A
Page 19: ...Chapter 2 Parts List c v 3 2 1 Parts and Accessories 2 2 1CONTENTS 1 2 1 RID 1OA...
Page 63: ...c Chapter 6 Initial Performance Test 1 6 1 System Performance Check 6 2 6 1 RID 1OA...
Page 102: ...Chapter I 1 Specifications 11 1 RID 10A Specifications 11 2 I CONTENTS I 11 1 RID 1OA...
Page 105: ...Parts I 12 1 Consumablesand Repair Parts 12 2 12 2 Optional Parts 12 4 12 1 RID 1OA...
Page 109: ...Chapter 13 Reference I i L 13 1 Solvent Characteristics 13 2 I 13 1 RID 1OA...