Chapter 4 Operating Panel
4-37
Setting Movement Targets
Specifies the target test force after preloading is complete.
[Setting Procedure]
[Setting Value]
Enters the target test force.
Parameter
Default Value
Setting Range
Target
0.002% of rating
0.002% of rating
to 100% of rating value
This setting is used to apply a preload to specimens before testing or to remove slack from
specimens.
Preload
Target
Pre.Load
>Target
10.00 N
SYSTEM