Chapter 4 Operating Panel
4-30
Switching Breakpoint Peak Level Function ON or OFF
Switches the breakpoint peak level function ON/OFF.
[Setting Procedure]
[Setting Value]
•
ON
Enables breakpoint peak level function. (default setting)
•
OFF
Disables breakpoint peak level function.
Setting the Breakpoint Peak Level Value
Specifies the breakpoint peak level.
[Setting Procedure]
[Setting Value]
Enters the breakpoint peak level.
Parameter
Default Value
Setting Range
Peak Level
90%
0.1% to 99.9%
The breakpoint peak level function determines the breakpoint at the point where the test
force extends a specified percent beyond the maximum test force setting value.
Break Setting
Peak Level
Switches function
ON or OFF
Break Setup
>PeakLevel Setup
>ON/OFF Setup
ON
METHOD
Break Setting
Peak Level
Peak Level
Break Setup
>PeakLevel Setup
>PeakLevel Value
50.00 %
METHOD
Summary of Contents for AGS-100NX
Page 2: ...This page is intentionally left blank...
Page 16: ...Before Using the System xiv This page is intentionally left blank...
Page 20: ...Contents xviii This page is intentionally left blank...
Page 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Page 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Page 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Page 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Page 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Page 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Page 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Page 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Page 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...