![Shimadzu AGS-100NX Instruction Manual Download Page 152](http://html1.mh-extra.com/html/shimadzu/ags-100nx/ags-100nx_instruction-manual_1256925152.webp)
Chapter 11 Troubleshooting
11-6
Alarm Code
70
Alarm Display
DataLety Err.
Description
Dataletty system does not enter ready mode.
Corrective
Measures
Confirm whether the Dataletty unit is online.
Alarm Code
9999
Alarm Display
System Err.
Description
An error other than those above occurred.
Corrective
Measures
Switch the AGS-X power OFF, then switch it ON again after about 10 seconds.
If the error continues after repeated attempts, it could be a hardware failure.
Contact your Shimadzu service representative.
In addition to system errors, if an error code other than indicated above is displayed, there
may be a hardware failure. Contact your Shimadzu service representative.
11.3 Other Problems
This section describes examples of problems that do not result in an alarm message.
1. Power Supply Problems
Problem
Cannot switch the instrument power ON.
Corrective
Measures
Confirm whether a circuit breaker has tripped on the power supply panel.
Confirm whether the power cord is disconnected.
Confirm whether the input voltage mode switch is set properly.
Confirm whether a fuse has failed.
If replacing fuses, contact your Shimadzu service representative.
Problem
Fuses fail frequently.
Corrective
Measures
Confirm whether the input voltage mode switch is set properly.
Confirm whether the fuses current rating is correct.
(100 V/120 V
→
10 A, 220 V/240 V
→
5 A)
Confirm whether a time-delayed type of fuse is being used. (A "fast-acting" fuse
cannot be used.)
If replacing fuses, contact your Shimadzu service representative.
Problem
A ground fault circuit interrupter was tripped on the power supply panel.
Corrective
Measures
There may be an electrical leak within the AGS-X unit. Immediately stop using
the system and contact your Shimadzu service representative.
Summary of Contents for AGS-100NX
Page 2: ...This page is intentionally left blank...
Page 16: ...Before Using the System xiv This page is intentionally left blank...
Page 20: ...Contents xviii This page is intentionally left blank...
Page 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Page 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Page 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Page 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Page 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Page 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Page 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Page 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Page 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...