Chapter 10 Maintenance and Inspection
10-3
The combined distance of (A) and (B) is about 4 mm. Since there is a return error of
about 1 mm for the free position, each value is between 1 mm and 3 mm.
If (A) or (B) is too small, vibration during testing could accidentally trigger the limit switch.
If the return error exceeds the rated value, contact your service representative.
•
Confirm that limits can be released by squeezing the slide limit handle.
•
Without releasing the slide limit, grip the body of the slide limit and try to pull the slide
limit upward. Confirm that the slide limit does not slip with respect to the limit bar.
•
Similarly, try to push the lower slide limit downward to confirm that the slide limit does
not slip with respect to the limit bar.
If a slide limit slips, contact your service representative.
Freely Movable Position
(A) Upper Activation Point
(B) Lower Activation Point
Summary of Contents for AGS-100NX
Page 2: ...This page is intentionally left blank...
Page 16: ...Before Using the System xiv This page is intentionally left blank...
Page 20: ...Contents xviii This page is intentionally left blank...
Page 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Page 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Page 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Page 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Page 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Page 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Page 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Page 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Page 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...