Chapter 4 Operating Panel
4-27
11. Break Detection
Setting the Start Position for Break Detection
Specifies the test force value to reach before starting specimen break detection.
[Setting Procedure]
[Setting Value]
Enters the test force value to reach before starting specimen break detection.
Parameter
Default Value
Setting Range
Start
7% of rating
0.035% to 7% of rating
Setting the Action After Break Detection
Specifies the crosshead action after detecting a break.
[Setting Procedure]
[Setting Value]
•
Stop
Crosshead is stopped after detecting break. (default setting)
•
Return
The crosshead returns quickly to its origin point after detecting
break.
Break Setting
Action
Start
Break Setup
>Start Point/Act
>Start Point
10.00 N
METHOD
Break Setting
Action
Action
Break Setup
>Start Point/Act
>Action
Stop
METHOD
Summary of Contents for AGS-100NX
Page 2: ...This page is intentionally left blank...
Page 16: ...Before Using the System xiv This page is intentionally left blank...
Page 20: ...Contents xviii This page is intentionally left blank...
Page 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Page 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Page 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Page 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Page 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Page 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Page 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Page 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Page 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...