Chapter 11 Troubleshooting
11-3
2. Alarm Code List
Alarm Code
1
Alarm Display
EMG Switch On
Description
The emergency stop button is pressed.
Corrective
Measures
After confirming safety, release the emergency stop button. To release the
emergency stop button, turn the movable portion in the direction indicated by
arrows or pull it up.
Alarm Code
2
Alarm Display
Cross-Head Limit
Description
Crosshead limit was activated.
Corrective
Measures
Temporarily loosen the crosshead limit collar and move the limit bar back to its
freed position. Use jog operations to move the crosshead to a safe position.
Alarm Code
10
Alarm Display
Servo Off
Description
Crosshead movement was attempted with the servo OFF.
Corrective
Measures
Confirm whether the emergency stop button or a crosshead limit is active.
Alarm Code
11
Alarm Display
Servo ALM
Description
An error has occurred in the servo amplifier.
Corrective
Measures
The servo motor could be overloaded or the power supply voltage may have
dropped. Switch the AGS-X power OFF and switch it ON again after about 10
seconds.
If this does not resolve the problem, there may be a hardware failure. Contact
your Shimadzu representative.
Alarm Code
12
Alarm Display
USB-Func (PC) Err.
Description
An error occurred during USB communications with the computer.
Corrective
Measures
Switch the AGS-X power OFF, then switch it ON again after about 10 seconds.
Also close and restart the materials testing software on the computer.
Alarm Code
16
Alarm Display
Inter Lock ALM
Description
The interlock for the fragment containment cover was activated.
Corrective
Measures
If this alarm occurred when starting a test
The fragment containment cover is open. Close the fragment containment
cover when starting a test. Also, the fragment containment cover must be
opened and closed after finishing one test before the next test can be
started.
If this alarm occurred during a test
The fragment containment cover was opened during the test. Do not open
the containment cover during tests.
Alarm Code
18
Alarm Display
PC Call Err.
Description
An error has occurred in the computer link function.
Corrective
Measures
Switch the AGS-X power OFF, then switch it ON again after about 10 seconds.
Also close and restart the materials testing software on the computer.
Alarm Code
20
Alarm Display
Over Rating ALM
Description
Test force has ex102% of rated load cell capacity.
Corrective
Measures
Use load cells within their rated capacity. Jig weight should be within
20% of
the load cell rating and zero the test force with the jig attached.
Summary of Contents for AGS-100NX
Page 2: ...This page is intentionally left blank...
Page 16: ...Before Using the System xiv This page is intentionally left blank...
Page 20: ...Contents xviii This page is intentionally left blank...
Page 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Page 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Page 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Page 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Page 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Page 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Page 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Page 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Page 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...