Chapter 10 Maintenance and Inspection
10-2
10.1 Periodic Inspections
Shimadzu offers periodic inspection services to ensure the testing machine functions reliably
for a long time. To ensure traceability, it is recommended that the testing machine be inspected
and serviced about once a year.
This service is performed on-site by Shimadzu trained service engineers using documents
provided by Shimadzu.
Shimadzu also offers validation services according to official standards as well. For more
information, contact your Shimadzu sales or service representative.
Shimadzu cannot warrant products repaired, adjusted, or otherwise serviced by a company
other than Shimadzu or its designated agents.
10.2 Inspection of Safety Equipment
If the system does not function properly during the following inspection procedures, contact
Shimadzu or its service representative. Shimadzu cannot warrant products repaired, adjusted,
or otherwise serviced by a company other than Shimadzu or its designated agents.
1. Emergency Stop Button
1) While pressing a jog button to move the crosshead, press the emergency stop button.
2) The crosshead should stop and not restart, even if the jog button is held down.
3) Reset the emergency stop button by turning the moveable portion in the direction
indicated by arrows.
2. Crosshead Stroke Limit Switches
1) While pressing a jog button to move the crosshead, push up on the upper slide limit and
hold it there.
2) The crosshead should stop and not restart, even if the jog button is held down.
3) Stop pushing up on the upper slide limit and let the limit bar move freely. In this state,
press a jog button to confirm that the crosshead can be moved.
4) While pressing a jog button to move the crosshead, push down on the lower slide limit
and hold it there.
5) The crosshead should stop and not restart, even if the jog button is held down.
6) Stop pushing down on the lower slide limit and let the limit bar move freely. In this state,
press a jog button to confirm that the crosshead can be moved.
This completes the inspection procedure. A diagram of the limit switch activation points is
shown on the next page.
Summary of Contents for AGS-100NX
Page 2: ...This page is intentionally left blank...
Page 16: ...Before Using the System xiv This page is intentionally left blank...
Page 20: ...Contents xviii This page is intentionally left blank...
Page 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Page 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Page 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Page 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Page 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Page 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Page 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Page 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Page 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...