Chapter 4 Operating Panel
4-24
Setting the Hold Period
Specifies how long to hold the crosshead when it reaches a maximum or minimum point.
[Setting Procedure]
[Setting Value]
Parameter
Default Value
Setting Range
Hold Time
0 sec
0 sec to 99999 sec
Setting the Hold Sensor
Specifies which input to keep constant during hold mode.
[Setting Procedure]
[Setting Value]
•
Stroke
Holds the stroke value constant. (default setting)
•
Force/Stress
Holds the test force or stress value constant.
If the hold period for single or cycle tests is set to a second or more, or the maximum point
action for cycle tests is set to [Manual], then the hold sensor must be selected.
Max. Point Action
Hold Period
Max.Action
>Hold Time
0 sec
METHOD
Max. Point Action
Hold Sensor
Max.Action
>Hold Sensor
Force
METHOD
Summary of Contents for AGS-100NX
Page 2: ...This page is intentionally left blank...
Page 16: ...Before Using the System xiv This page is intentionally left blank...
Page 20: ...Contents xviii This page is intentionally left blank...
Page 21: ...1 1 Chapter 1 Overview CONTENTS 1 1 Introduction 1 2 1 2 AGS X Unit Specifications 1 3...
Page 38: ...Chapte 2 Part List Specifications 2 12 This page is intentionally left blank...
Page 56: ...Chapter 3 Basic Instructions 3 18 This page is intentionally left blank...
Page 118: ...Chapter 5 Performing Tensile Tests 5 12 This page is intentionally left blank...
Page 126: ...Chapter 6 Performing Compression and Bending Tests 6 8 This page is intentionally left blank...
Page 130: ...Chapter 7 Method Files 7 4 This page is intentionally left blank...
Page 134: ...Chapter 8 Control Input Output 8 4 This page is intentionally left blank...
Page 142: ...Chapter 9 Sensor Input Output 9 8 This page is intentionally left blank...
Page 146: ...Chapter 10 Maintenance and Inspection 10 4 This page is intentionally left blank...