Plasma Gas Analyzer
71
02001001A_9
The default value is normally just what the system needs and, under normal conditions, this
value should not be changed.
From this menu, by selecting "F3" for next sub-menu, the following menu is brought up on the
display.
Figure 51 -
BACKGROUND GAS SELECTION MENU
This menu is used for display and selection of the background gas type. The background
gas may be helium or argon.
The
"F1"
key toggles between argon and helium.
NOTE : When selecting helium, the sample
flow will be set at 75 cc/min. Selecting back argon will set the sample flow set point at 75
cc/min. You may change later the sample flow set point in the configuration menu. See
User’s Manual
. After the selection is done, use
“F3”
to go in the next menu or
“
F4”
to quit.
Figure 52 below shows the next menu.
Figure 52 - TEMPERATURE COEFFICIENT MENU
The temperature coefficients are defined during bench testing at the factory. These values
are used to do temperature compensation for ambient temperature change if required. You
should not change this value unless you know how to proceed for temperature bench
marking. A value of 1.000 cancels the temperature compensation.
<<<<BACKGROUND GAS SELECTION>>>>
ACTUAL GAS : HELIUM
Use F1 key for selection
F3 : NEXT
F4 : QUIT
<<<<
TEMPERATURE COEFFICIENTS>>>>
F1: OFFSET
F2: GAIN
F3: NEXT
F4: QUIT
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...