Plasma Gas Analyzer
16
02001001A_9
4.2
Electrical Installation
4.2.1 Electrical Safety
WARNING
Ensure that the electrical installation of the analyzer conforms with all
applicable local and national electrical safety requirements.
WARNING
Obey the safety instructions given below when you install the analyzer;
if you do not, the analyzer warranty may be invalidated, the analyzer
may not operate correctly, or it may be damaged.
The electrical supply coupler or plug must be easily accessible for
disconnection from the electrical supply.
Ensure that your electrical supply can provide the necessary maximum
power consumption: refer to Section 10
All signal and electrical supply cables must be rated for temperatures
of 70 °C or higher.
Ensure that the cables that you connect to the analyzer are routed so that
they do not present a trip hazard.
When you carry out insulation testing, disconnect all cables from the
analyzer.
Ensure the analyzer is provided with a sound earth connection via the
electrical supply plug (Figure 2 Item 3).
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...