Plasma Gas Analyzer
77
02001001A_9
Status
Acceptable output
Description
STATUS
ST
fs
fz
fw
Three possible combinations here:
ST means there are no bad calibration errors,
fs means the analyzer failed to span
fz means the analyzer failed to zero
In the case that there has been a bad span
and a bad zero, the last bad calibration is
shown. When that one is done right then, the
other bad calibration is shown. Both are
cleared when both calibrations are done right
or unit is reset.
fw is used whenever a different kind of failure
is triggered by the analyzer that requires
operator attention by reading a message on
the analyzer such as the plasma being off.
This flag should reset itself if the error
condition goes away (go back to ST)
CONCENTRATION
37.90 ppm
The concentration number must have two
decimal places, for example:
1.01
21.10
501.23
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...