5.8.2 Temperature Coefficient
This menu gives the possibility to enter coefficient that will be used for system temperature
variation. Using these factors, the software will do a temperature compensation to minimize
temperature drift. There are two factors, one for the offset and another for the gain. The
temperature affects the baseline of the instrument and the sensitivity. These factors are
entered at the factory during temperature drift tests. Most of the time these factors will be set to
1.000 and should not be changed. If you suspect there is an error call Servomex to get the exact
procedure to perform the temperature tests.
Press F1 to access the
TEMPERATURE OFFSET MENU
.
Figure 30 - TEMPERATURE COEFFICIENT
Minimum: 0
Maximum: 2
Resolution: 0.001
If you have to change the offset compensation factor, enter the value with the numeric keypad
and then, press
“E”
to validate the new value.
Press
“F2”
to access the temperature gain menu.
<<<TEMPERATURE OFFSET>>>
Offset:1.000
Actual:1.000
F4:RET
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...