Plasma Gas Analyzer
62
02001001A_9
Figure 47 - AUTOMATIC CALIBRATION TIMERS
Minimum: 0
Maximum: 90000.0 (between calibration) and 25000.0 (on calibration gas)
Set the
“TIME BETWEEN CALIBRATIONS”
by pressing
“
F1
”
and entering the value in hours
followed by the E key. Press F2 and enter a value in minutes followed by the E key to set the
“TIME ON CAL. GAS”.
To set the
“TRANSFER DELAY”
, advance in the hidden menu by pressing
“F3”
until the
following page is displayed:
Figure 48 -
ANALOG OUTPUT HOLD TO TRACK TRANSFER DELAY
Minimum: 0
Maximum: 90000.0 (between calibration) and 25000.0 (on calibration gas)
<<<AUTO CALIBRATION PARAMETERS>>>
F1:Time between calibration: 168.0 Hrs
F2:Time on cal. gas:10.0 Min
F3:NEXT
F4:RET
<<<ANALOG OUTPUT>>>
Hold to track transfer delay:120.0 Sec
New Value:0.0
F4:RET
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...