Figure 18 - CONFIGURATION MENU - PAGE 2
5.3.4 mA failure mode
mA failure mode
“
F1
”
: in the case of an “underscale”, "over scale", “low flow” or "plasma off",
this option will set the 4-20 mA output :
“off “(always remain between 0 and 20 mA, even if an
error occurs),
“low” (below 4 mA), or “high” (higher than 20 mA).
If you have any options, you will find them described in section 4.2.5 of this manual.
You exit the configuration section by pressing
F4
. The analyzer will display the first page of the
CONFIGURATION MENU
. Pressing
F4
again brings up the
MAIN MENU
.
This is the
standard configuration. Other menus may appear if your analyzer has other optional
features (serial port, fully automatic calibration, digital I/O, etc.).
NOTE: When you select the automatic ranging feature, the scale selection is based on
analogue to digital count value or ppm value, based on which one reaches the maximum value
for the scale in use.
<<CONFIGURATION MENU>> Page 2
F1:mA failure mode (high/low/off):HIGH
No other options available
F4:RET
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...