Plasma Gas Analyzer
23
02001001A_9
The communication parameters of the analyzer are:
Baud rate:
9600
Parity bits:
none
Data bits:
8
Stop bits:
1
The serial port on the rear of the analyzer (Figure 2 item 6) is an RS232 ± 5.5 V 9-way
'D' type connector.
Pin(s)
Use
1
Not used
2
Rx (to the analyzer)
3
Tx (from the analyzer)
4
Not used
5
0 V
6, 7, 8, 9
Not used
1
Figure 11 - RS232 CONNECTION PIN DETAILS
Standard RS-232 Interface - Transmission protocol
Note: see Appendix 4 for details of the transmission protocol for the Enhanced Interface
for use with External Valve Box
The different parameters are transmitted in the following order for the standard
interface:
ppm value sign, ppm value,
TAB,
flow,
TAB,
flow counts,
TAB,
cell counts
TAB,
status
and range,
TAB,
checksum,
CR
.
Each parameter is formatted as followed:
PPM value sign
: 1 byte equal to “+” or “-” in ASCII, depending on the
polarity of the ppm value;
PPM value
: 6 bytes in ASCII (3 digits before the dot and two after);
Flow
: 6 bytes in ASCII (3 digits before the dot and two after);
Flow counts
: 8 bytes in ASCII;
Cell counts
: 8 bytes in ASCII;
Status and range
: bitwise (logical on 8 bits);
1 2 3 4 5
6 7 8 9
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...