Plasma Gas Analyzer
24
02001001A_9
bit 7: alarm 2
1 if alarm 2 is ON
bit 6: alarm 1
1 if alarm 1 is ON
bit 5: low flow
1 if there is a low flow error
bit 4: plasma status
1 if there is a plasma off error
bit 3: system status
1 if there is a low flow, plasma off, underscale
or overscale error
bit 2, and 0: range in use
001 = range 1 (ex: 0-1 or 0-10 ppm)
010 = range 2 (ex: 0-10 or 0-50 ppm)
100 = range 3 (ex: 0-100 ppm)
Each value is separated by a TAB character. The end of one transmission
(packet or frame) is indicated with a carriage return.
The checksum is the arithmetic sum of every transmitted byte (excluding TAB
bytes). The checksum may be used to verify the data integrity.
Example
: the ppm value is 40.1, the flow is 75.0 ml min
-1
(75 cc/min), the flow
counts are 8388600, the cell counts are 190011, the range used is the first one
(0-1 ppm), there are no alarm, but there is a low flow error. The data sent in
ASCII is:
Data
Check sum
“+”, “NUL”, “4”, “0”, “.”, “1”, “0”,
286
“TAB”,
“NUL”, “7”, “5”, “.”, “0”, “0”,
250
“TAB”,
“NUL”, “8”, “3”, “8”, “8”, “6”, “0”, “0”,
369
“TAB”,
“NUL”, “NUL”, “1”, “9”, “0”, “0”, “1”, “1”,
300
“TAB”,
“)”,
41
“TAB”,
Total
: 1246
1”, “2”, “4”, “6”,
“CR”
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...