Plasma Gas Analyzer
79
02001001A_9
The following table lists the commands that the external Valve Box can send to the analyzer.
Command
Description
as<enter>
Immediately starts an auto-span
az<enter>
Immediately starts an auto-zero
al<enter>
Local Regime. Valve Box is under analyzer control. In this mode the
analyzer can be spanned and zero from the front panel pf the
instrument, and it can control valves in the Valve Box.
When analyzer is powered up, it defaults to local regime mode. The
Valve Box on receiving the string output through the RS-232 port will
send the “ar” command to set the instrument in remote mode if
needed.
ar<enter>
Remote Regime. Valve Box is under control of the PLC. The
analyzer has the span and zero functions disabled while in Remote
Regime, only the Valve Box can initiate span and zero through the
RS-232 port
The following table gives the setting requirements of RS-232 port of the Valve Box.
Parameter
Setting
Baud
9600
Byte
8 bits
Parity
None
Stop Bits
1
Message Interval
2 seconds
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...