Plasma Gas Analyzer
33
02001001A_9
Figure 21- DISPLAYED WHEN ENTERING SPAN VALUE
The analyzer will automatically select the appropriate range based on the span value entered.
When you have entered a value and pressed the
"E"
key (E for enter), the processor checks the
format, evaluates it and validates it. If everything is OK, your value will be redisplayed in the
actual value field. If not, the value is rejected, the input field is cleared and the actual value is left
unchanged.
If you made an error when entering your value, just press enter and start over.
Once you have entered the zero and the span values, select
“F3”
to get the
CALIBRATION
SUB-MENU.
The following figure shows this sub-menu:
Figure 22 - CALIBRATION SUB MENU
<<CALIBRATION FACTORS CALCULATION>>
Measure: 9 PPM Sample Flow: 25 CC
F1:Zero F2:Span F4:RET
Calibration status:
<<CALIBRATION MENU>>
Input SPAN value:0.0
Actual value:0.0
Value must be in PPM F4:RET
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...