Plasma Gas Analyzer
70
02001001A_9
3.5
Configuration and operation
3.5.1 Introduction
All configuration, calibration, diagnostic and run menus described in the standard User
Manual also apply to the argon/helium version of the SERVOPRO PLASMA series analyzer.
There is a supplementary
"HIDDEN"
menu to select the starting count value and the
background operating gas.
3.5.2 HIDDEN menu access
To access the
HIDDEN MENU
, go to the main menu and enter the secret code by pressing
“1”, “2”, “3”,
and
“E”
consecutively on the keyboard. This brings up the hidden menu see
Figure 50 -
OPERATING PARAMETERS MENU
The starting count value is the threshold value where the analyzer switches from starting
mode to normal mode.
If the cell count value (value from the analog to digital converter; this value may be observed
under
DIAGNOSTIC MENU
) is under the starting count value, a higher power is applied to
the cell by the plasma generator. Once the plasma is started, the cell count will be higher
than the starting count value and the power will come back to normal. If, for any reason, the
cell count value with the normal power applied to the cell is too close or oscillates around the
starting count value, the analyzer will go back and forth between starting and normal mode.
To get rid of such a situation, you should enter a new (lower) threshold value.
To enter a new value, just use the numerical key on the keypad followed by pressing the
"E"
key. The value must be between 0 and 1,000,000 counts.
<<<<OPERATING PARAMETERS>>>>
ACTUAL STARTING COUNT VALUE :
500
INPUT NEW VALUE :
F3 : NEXT
F4 : QUIT
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...