Plasma Gas Analyzer
63
02001001A_9
Enter a realistic value based on your system configuration, typically around 5 minutes. You
should make sure that this value is big enough to allow the sample gas to purge the span gas in
the system.
When the proper timing parameters are entered, automatic calibration can be turned on from
the calibration selection mode menu Figure 42
Toggle the calibrat
ion mode until “
AUTOMATIC”
is selected by pressing F1 in the calibration
selection mode menu. The calibration timer “
TIME BETWEEN CALIBRATIONS”
begins to
count down from the moment you put the analyzer in automatic calibration mode. You may track
the automatic calibration process and monitor both timers
“TIME BETWEEN CALIBRATIONS”
and
“TIME ON CAL. GAS”
from the diagnostic menu. The value displayed next to
“ELAPSED TIME SINCE LAST CAL.”
is the number of hours since the analyzer was put in
automatic calibration mode. The value displayed next to
“MINUTES ON CAL. GAS”
is the
number of minutes the analyzer has spent on calibration gas (zero or span) since the zero or
span gas was selected. “0.0” is displayed if the analyzer is not currently performing a
calibration.
Furthermore, in the run menu, the gas currently selected is displayed next to
“Gas”.
Note that
during an automatic calibration sequence, the status contact remains OPEN.
When the analyzer is in the
RUN
menu during an automatic calibration, the word
“AUTOCALIB”
will be displayed instead of the measure in PPM.
NOTE: In order to get accurate readings when using auto-calibration mode, the user must enter
proper timing values in accordance with the sampling system. Longer purging time are better
than shorter ones.
NOTE: Only one alarm can be used with the auto-calibration option.
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...