Plasma Gas Analyzer
58
02001001A_9
1.2
System Operation
Calibration of the analyzer may be done manually or automatically at a time interval defined by
the user. In both cases calibration gases are selected with SOV1, SOV2 and SOV3. This
section explains how to execute a calibration manually and automatically.
NOTE: Zero and span gas values must be set prior to any calibration with a span value
bigger than the zero value.
1.3
Entering Zero and span gas values
From the main menu, press
“
F2”
to enter the calibration menus of the analyzer.
proceed to the
CALIBRATION MAIN MENU
by pressing
“F3” and follow standard steps as
described in section 5.4
1.4
Manual Calibration
From the main menu, press
“F2”
to enter the calibration menus of the analyzer.
Figure 42 - CALIBRATION MODE SELECTION MENU
Select the
“MANUAL”
mode if you want to perform a manual calibration toggle the calibration
modes by pressing
“F1”
until
“MANUAL”
is displayed. Then, toggle between gases by
pressing
“F2”
until
“ZERO”
is displayed (SOV2 is energized).
<<CALIBRATION MENU>>
F1: Calibration Mode: MANUAL
F2: ACTUAL GAS: SAMPLE
F3: NEXT
F4: RET
Summary of Contents for SERVOPRO Plasma
Page 1: ...Part Number 02001001A Revision 9 Language UK English SERVOPRO Plasma OPERATOR MANUAL...
Page 2: ...This page intentionally blank...
Page 10: ...Plasma Gas Analyzer 8 02001001A_9 Figure 3 INTERNAL SCHEMATIC...
Page 16: ...Plasma Gas Analyzer 14 02001001A_9 Figure 4 DIMENSIONAL OUTLINE DRAWING...
Page 58: ...Plasma Gas Analyzer 56 02001001A_9 APPENDIX 1 AUTOMATIC CALIBRATION OPTION...
Page 66: ...Plasma Gas Analyzer 64 02001001A_9 Figure 49 SERVOPRO PLASMA WITH AUTOMATIC CALIBRATION...
Page 67: ...Plasma Gas Analyzer 65 02001001A_9 APPENDIX 2 TRACE NITROGEN IN HELIUM VERSION...
Page 70: ...Plasma Gas Analyzer 68 02001001A_9 APPENDIX 3 DUAL BACKGROUND ANALYZERS...
Page 76: ...Plasma Gas Analyzer 74 02001001A_9 APPENDIX 4 INTERFACE FOR EXTERNAL VALVE BOX...