Practical Example: Diameter Determination
Menu code settings used in the example:
Function
Code
Diameter determination
2 1 7
Automatic output of
length and density
7 1 2
With data ID code
7 2 2
Application:
Determination of the diameter of wires and metal filaments
(e.g., for tungsten or molybdenum)
Step/Key ID/Key
Readout
Data Output
Activate the parameter mode
rHo
by pressing
i
>2 seconds
8.000 °
mi
Enter density for filament
(in this case, tungsten):
19.25
19.25
mg
Store value in the non-volatile
memory by pressing STO -
F
Zero the balance using
t
0.000 mg
Place tungsten filament on pan; + 59.423 mg
close the draft shield
Enter filament length, e.g.,
200
(mm); store using
200
L
+
200 mm
STO -
F
+
0.140 °
mi
RhoG
+
19.25 g /
Print diameter by pressing
p
+
0.140 °
mi
Dia
+
0.140 mm
Taggle to weight readout with
w
+ 59.423 mg
The data ID codes
mean:
L
+
200 mm
length of filament
RhoG
+
19.25 g /
density of the object weighed (e.g., filament)
Dia
+
0.140 mm
calculated diameter
* = factory setting
As an alternative, you can select
this code:
Data output – manual mode 7 1 1*
3–43