The calculated value is locked into the display once the sample residue has been
stored. To “unlock” this readout, press RE S v -
F
or select a different memory
location using the numeric keys and
F
. Afterwards, you can proceed with the
next tare weighing.
A stored value of a backweighed sample can be overwritten with a current value
anytime after the corresponding memory location has been selected. To do this,
press
c
. A backweighing operation is done in reference to the original sample
weight (= 100%).
Once you have completed processing or treatment of a series of samples, you can
select another memory location using the numeric keys. Therefore, you can
perform backweighing in any sequence (known as “random backweighing”). In
the consecutive individual weighing mode, the next memory location is
automatically selected using the RE S v -
F
key after backweighing (code 3 6 2).
* including the Signatories of the Agreement on the European Economic Area
Important Note for Verified Balances Approved for Use as
Legal Measuring Instruments in the EU:*
You may not use the calculated value when the balance is being used
as a legal measuring instrument. The display symbol “R2” flashes to identify non-
verified weights (not legal for trade).
3–20
Summary of Contents for MC 210 S
Page 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Page 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Page 46: ......
Page 78: ......
Page 160: ......
Page 166: ...5 6 Z 14 755 320 34 10 93...
Page 170: ......
Page 174: ...6 4...
Page 178: ......
Page 179: ......