Air Density Determination
For highly precise analyses, the balance offers you a function with which
fluctuations in the air density can be determined.
!
The air density at the place you set up your balance can be determined
within the range of 1.0 kg/m
3
and 1.4 kg/m
3.
For determining the air density, use the special weight set which is available
as an accessory (see also part 5, “Accessories”). This weight set consists of one
steel and one aluminum weight. Their exact mass values are specified on
a certificate, which is included with the set. By a differential weighing method, the
air density is determined using the two weights.
To access this function, set the appropriate code in the balance operating menu:
Air density determination
Code
Accessible
1 16
1
Access denied
1 16
2*
Use the
c
key at any time to stop air density determination.
* = factory setting
3–9