Step/Key ID/Key
Readout
Data Output
Enter memory location/
container ID; e.g.,
7
,
7
and press
F
to confirm
Zero the balance using
t
0.000 mg
Place container on pan;
+ 376.307 mg
store weight using
●
-
F
0.000 mg
Net
Add prepared sample to
+1095.335 mg
Net
container; store weight
using STO -
F
+1435.642 mg
Treat sample (e.g., incinerate)
Place container with treated
sample on pan
+ 991.571 mg
Backweigh using RE S -
F
+ 615.264 mg
Net
Num
7
Date
19-Jul-95
Time
10:34:54
Net
+ 1059.355 mg
Res
+
615.264 mg
Res
+
58.08 %
Toggle to readout of 2nd
+
58.08 %
g
Net
calculated value:
w
Repeat procedure for the next
8
memory location: RE S v -
F
, or
to delete data stored in the
memory location for this
sample, press
c
> 2 sec.
The data ID codes
mean:
Num
7
memory location ID number
Date
9-Jul-95
date of backweighing
Time
10:34:54
time of backweighing
Net
+ 1059.355 mg
net initial sample weight
Res
+
615.264 mg
net residue (backweighed sample) in wt. unit
Res
+
58.08 %
residue in percent
* = factory setting
3–22