Weighing Sequence
Individual or Combined Weighing Mode
Individual Weighing Mode
In the
individual weighing mode,
the weights are stored in the following sequence:
– Tare weighing – weighing-in – backweighing
– If necessary, selection of an available memory location
– Tare weighing – weighing-in – backweighing...
The data of a sample in a memory location are overwritten until a new memory
location is selected.
In the
consecutive individual weighing mode,
the weights are stored in this sequence:
– Tare weighing – weighing-in – backweighing
– Automatic selection of the next memory location
– Tare weighing – weighing-in – backweighing...
Beginning at a specific memory location, the memory locations that follow
are selected automatically using the
F
key. It is also possible to switch to a new
memory location using the numeric keys and the
F
key.
Combined Weighing Mode
In the
combined weighing mode,
the values are stored in the following sequence:
– Tare weighing – weighing-in – tare weighing – weighing-in...
Backweighing is done only after all of these operations have been completed.
2
1
1
2
3–15
Summary of Contents for MC 210 S
Page 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Page 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Page 46: ......
Page 78: ......
Page 160: ......
Page 166: ...5 6 Z 14 755 320 34 10 93...
Page 170: ......
Page 174: ...6 4...
Page 178: ......
Page 179: ......