Practical Example: Checking Net Weights
Menu code settings used in the example:
Function
Code
Checking net weights
2 2 3
Control by port lines for the
YRD10 Z checkweighing display 8 7 1
Tol target weight
7 1 2
With data ID code
7 2 2
Application:
Weighing-in the same amounts of samples
Step/Key ID/Key
Readout
Data Output
c
; place sample container
on pan;
t
0.000 mg
Weigh in required
sample amount
+ 493.110 mg
Store: STO -
f
1.0 %
g
Lim
+
1.0 %
Setp
+
493.110 mg
+ 493.110 mg
Remove con ideal
sample; place another
container on pan;
t
0.000 mg
Weigh in sample using
Checkweighing
the analog bar graph
indication
as a guide
+ 495.894 mg
p
+ 495.894 mg
N
+
495.894 mg
Continue as directed above
The data ID codes
mean:
Lim
+
1.0 %
over/under limits in percent
Setp
+
493.110 mg
target weight stored
N
+
495.894 mg
net sample weight
As an alternative, you can select
this code:
Target weight only
7 1 3
3–52