Data Output Formats for the “Draft Shield Status”
These depend on the particular menu code setting:
code 7 2 1 “without data ID code” = data output with 16 characters,
or code 7 2 2 “with data ID code” = data output with 22 characters
Data Output Format with 16 Characters
Data Output Format with ID Code (22 Characters)
With this format, the ID code “Stat” is output in the 1st through the 4th places.
* = space
The draft shield status consists of control codes and the turning angle of the
draft shield. The control codes are identified as
xx
, and the turning angle
is designated as
yyy
.
xx = Control Codes for the Draft Shield
The decimal values of bits 0 through 6 are added for the following control
codes and output as a decimal number:
Decimal
Binary Control
code
value
value
0
Bit 0 = 0
No control error
1
2
0
Bit 0 = 1
Control error
0
Bit 1 = 0
Draft shield motor “off”
2
2
1
Bit 1 = 1
Not implemented
0
Bit 2 = 0
Not implemented
4
2
2
Bit 2 = 1
Motorized draft shield, as far as it will turn
0
Bit 3 = 0
“Self-teaching” function off
8
2
3
Bit 3 = 1
“Self-teaching” function on
0
Bit 4 = 0
Open draft shield
1
)
16
2
4
Bit 4 = 1
Closed draft shield
1
)
0
Bit 5 = 0
Slow speed
1
)
32
2
5
Bit 5 = 1
Fast speed
1
)
0
Bit 6 = 0
Motorized draft shield operation
64
2
6
Bit 6 = 1
Manual draft shield operation
1
) = These control codes refer to the previous movements of the draft shield door.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S
t
a
t
* * * * * * * * W * 0 X X Y Y Y CR LF
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * * W * 0 X X Y Y Y CR LF
4–11
Summary of Contents for MC 210 S
Page 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Page 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Page 46: ......
Page 78: ......
Page 160: ......
Page 166: ...5 6 Z 14 755 320 34 10 93...
Page 170: ......
Page 174: ...6 4...
Page 178: ......
Page 179: ......