Specifications
for Verified Balances
with an EC Type Approval
Model
MC 210 S
MC 210 P
MC 410 S
-0CE
-0CE
-0CE
Type
KB BA 100
KB BA 100
KB BA 100
Accuracy class
k
k
k
Weighing range structure
SuperRange
PolyRange
SuperRange
Maximum capacity Max*
g
210
210
410
Scale interval d*
mg
0.01
0.01/0.02/0.05 0.1
Verification scale interval e*
mg
1
1
1
Minimum capacity Min*
mg
1
1
10
Tare range (subtractive)
%
–––––––––––––––
≤
105 of Max ––––––––––––––
Range of use according to
Directive No. 90/384/EEC*
g
0.001–210
0.001–210
0.001–410
Response time (average)
s
––––––––––––– 10 ––––––––––––––––– 2.5 –––
Adaptation to ambient conditions
and application requirements
s
––– By selection of 1 of 4 optimized filter levels ––
Display update (dep. on the filter level select.)
–––––––––––––––– 0.2–0.4 ––––––––––––––––
Allowable ambient operating temperature
°C
–––––––––––––––– +5...+40 ––––––––––––––––
Operating temperature range
°C
––––––––––––––– +10 ... +40 ––––––––––––––
Pan size
mm
–––––––––––––––––– Ø 90––––––––––––––––––
Weighing chamber approx., (Ø x H)
mm –––––––––––––––– 167 x 245 –––––––––––––––
Net weight, approx.
kg
–––––––––––––––––––12 –––––––––––––––––––
Power requirements
V~
TNG6 230 VAC or 115 VAC, –20%...+15%
Frequency
Hz
––––––––––––––––– 48–60 ––––––––––––––––
Power consumption approx.
VA
––––––––– maximum: 28; average: 16 ––––––––
Selectable weight units
g, mg
Selectable application programs
Mass unit conversion; “EUREKA” air buoyancy
correction progr.; diff. weighing and backweighing;
density determination; tare memory; net total;
weighing in percent; over/under checkweighing
(classification and filling); counting
Automatic zero tracking function
(can be turned off by menu code)
–––––––––––––––– Standard ––––––––––––––––
Built-in interface
RS-232 C-S/V24-V28, RS-423/V 10; 7-bit;
parity: even, odd; transmission rates:
150...19,200 baud; 1or 2 stop bits;
software/hardware handshake
Standard features/equipment supplied
Dust cover/power supply
x
x
x
Hanger for below-balance weighing
x
x
x
Self-calibrating function “isoCAL”
x
x
x
ISO/GLP recording capability/numeric keys
x
x
x
Numeric keys
x
x
x
* = EC Directive No. 90/384/EEC for non-automatic weighing instruments applicable in the
EC/EU Member States
5–4
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲
▲▲▲▲
▲
▲
▲
▲
▲
▲
Summary of Contents for MC 210 S
Page 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Page 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Page 46: ......
Page 78: ......
Page 160: ......
Page 166: ...5 6 Z 14 755 320 34 10 93...
Page 170: ......
Page 174: ...6 4...
Page 178: ......
Page 179: ......