Activating the Air Density Determination Function
Hold down the tare key for more than 2 seconds until
“
C.I.
” and “
CAL
” are displayed next to the
f
key.
To activate the
air density determination
function,
press
F
, identified by “v,” several times until “
A.d.
”
is displayed.
“
A.d.
” stands for “
air density.
”
Unload the weighing pan and tare the balance.
Starting Air Density Determination
To start air density determination, press the
f
key
when the display shows a zero readout. Afterwards,
a weight for steel “St” is displayed.
”Err 02”:
The effects of an external disturbance
can cause the error code “
Err 02
” to
be displayed briefly. If this occurs,
tare the balance and then press the
f
key again when the display shows
a zero readout.
Interrupt
procedure:
You can stop an air density
determination procedure
at any time:
– By pressing the
c
key so that
a previously stored air density
is not changed or
– By pressing
F
identified in the
display by “RES” to set the standard
air density 1.2 kg/m
3
3–11
Summary of Contents for MC 210 S
Page 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Page 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Page 46: ......
Page 78: ......
Page 160: ......
Page 166: ...5 6 Z 14 755 320 34 10 93...
Page 170: ......
Page 174: ...6 4...
Page 178: ......
Page 179: ......