Step/Key ID/Key
Readout
Data Output
Enter memory location ID no.,
e. g.,
10
and store using
●
-
F
10
Zero the balance using
t
0.000 mg
Place substrate on pan and
+1221.284 mg
store using STO -
F
+1221.284 mg
11
Remove substrate
0.000 mg
Place next substrate on pan and +1232.265 mg
store using STO -
F
+1232.265 mg
Remove substrate and weigh-in
12
further samples
Coat material
If necessary for a longer process,
use the
w
key to turn the
program off and back on again
Enter 1 st memory location ID
10
10
and store using
●
-
F
Zero the balance using
t
0.000 mg
Place coated material on pan; +1349.793 mg
backweigh using RE S -
F
+ 128.509 mg
Num
10
Date
9-Jul-95
Time
10:44:54
Net
+ 1221.284 mg
Remove material and
Res
+ 1349.793 mg
backweigh further samples
Diff
+
128.509 mg
using RE S v -
F
11
The data ID codes
mean:
Num
10
memory location ID number
Date
9-Jul-95
date of backweighing
Time
10:44:54
time of backweighing
Net
+ 1221.284 mg
net initial sample weight
Res
+ 1349.793 mg
net residue (backweighed sample) in wt. unit
Diff
+
128.509 mg
difference in weight unit
3–29
Summary of Contents for MC 210 S
Page 4: ...General View of the Balances MC 210 S MC 210 P and MC 410 S 0CE 1 0...
Page 6: ...General View of the Balances MC 5 0CE and SC 2 0CE 1 2 34 35 36 10 7 9 8 37...
Page 46: ......
Page 78: ......
Page 160: ......
Page 166: ...5 6 Z 14 755 320 34 10 93...
Page 170: ......
Page 174: ...6 4...
Page 178: ......
Page 179: ......