IDS701-CTRUCK   

Weighing Indicator 

 

 

 

 

User manual

 

 
 
 
 
 
 

 

 

 

Summary of Contents for IDS701-CTRUCK

Page 1: ...IDS701 CTRUCK Weighing Indicator User manual ...

Page 2: ...ate inspect and fix the the weighing indicator is prohibited by Non professional staff WARNING Pls make sure the weighing display well earthing WARNING The indicator is electrostatic sensitive device pls power off during electrical connections internal components touched by hand is prohibited and please take the anti static measure ...

Page 3: ... load cell 4 2 2 Connection of interface 4 3 Basic operation 6 3 1 Display and key instruction 6 3 2 Power on 8 3 3 Zero function 8 3 4 Print function 8 3 5 Re print function 8 4 Calibration 9 4 1 Enter and exit setting 9 4 2 Calibration steps 9 4 3 Parameter setting 12 5 Output data format 16 6 Maintenance 18 ...

Page 4: ...15 times sec load capacity it can connect 4 pcs 350Ω load cell at most weight unit kg Resolution 3000e Interval 1 2 5 10 20 50 Display 6 digits LED word height 20 3mm key ON OFF ZERO PRINT RE PRINT Interface RS232C Baud rate optional 1200 2400 4800 9600 Ambient temperature 10 40 Storage temperature 20 60 power AC220 ...

Page 5: ... four pcs 350Ωload cell at most both four and six wire load cell are ok To make it simple we use quick disconnect Or standard plug As belows Quick disconnect connection When use four wire load cell connect Exc to Sen connect Exc to Sen 2 2 Connection of interface RS232 communication interface for scoreboard use 3 cores quick connector ...

Page 6: ...ace connection RS232 communication interface for printer use DB9 definition 5 4 3 2 1 9 8 7 6 RS232 communication interface connection Pin function and definition as bellows DB9 joint Definition Function 2 TXD Sending data 3 RXD Receiving data 5 GND Ground interface ...

Page 7: ...IDS701 CTRUCK User Manual Page 6 of 21 3 Basic operation 3 1 Key and display IDS701 CTRUCK Weighing indicator display instruction LED display instruction Weighing data display ...

Page 8: ...f 21 Kg Weight unit kg Tare Display tare weight Net Display net weight Gross Display gross weight Display data keep still zero indicating zero weight Hi Over setting weight OK Within setting weight Lo Below setting weight Decimal point ...

Page 9: ... 2 seconds to power on or power off 3 2 Power on Power on and indicator perform self checking and go to weighing mode 3 3 Zero setting Within zero range press zero indicator weighing is cleared When Indicator is not stable zero is unworkable 3 4 Print function When the data is stable connection with printer it will be printed after press PRINT 3 5 RE Print function connection with printer press RE...

Page 10: ...Press RE PRINT still and then press ON OFF enter C08 C39 setting 2 Take out the sealing screw on the back of indicator then press down the span Press RE PRINT still and then press ON OFF enter C01 C39 setting The key functions in setting Enter Up Left Power switch exit setting 4 2 Step of calibration operation According to the second method which can enter setting menu C01 C39 ...

Page 11: ...ess or C03 C03 1 C03 5 Set graduation option 1 2 5 10 20 50 Select required graduation example graduation 5 C03 5 5 press press press or C04 0100 00 0100 00 Max capacity example max weighing 100kg 0100 00 6 press press press press C05 C05 0 C05 1 CAL 9 0000 00 Zero calibration option 0 non calibration zero 1 need calibration zero calibration zero please choose 1 and ensure scale is empty and stabl...

Page 12: ... example the weight is 80kg As bellows enter loading calibration count down over indicator shows loaded weight loading calibration finish If you want to set application Function parameter Press PRINT if you want to exit press TOTAL 8 press press press or C07 07 0 07 1 Default parameters setting option 0 non restore default parameters 1 restore default parameters Note after the above parameters set...

Page 13: ...tomatically 60 keep still within 60 min power off automatically Power saving setting C10 Power saving setting option 0 close power saving setting 3 keep still within 3 min stop display 5 keep still within 5 min stop display Hold function C11 Hold mode option 0 close hold function 1 Peak hold 2 Data Hold instruction Peak it shows the max data mainly application for materials testing such as tension...

Page 14: ...t date from left to right year month day Communication setting C18 Serial interface data output method option 0 Close serial interface data output 1 Continuous sending connect big display 2 print method connect printer 3 Command request method connect computer 4 PC continues to sending format connect computer C19 Baud rate option 0 1200 1 2400 2 4800 3 9600 Zero range C20 Manually zero range optio...

Page 15: ...set graduation 2 Automatic zero tracking range can not exceed manual zero setting range C23 Automatic zero tracking time option 0 close automatic zero tracking time 1 1 second 2 2 seconds 3 3 seconds Overload range C24 Overload range option 00 close overload range 01d 99d remark d is the setting graduation division Negative display C25 Negative display Option 0 9d 10 10 max capacity 20 20 max capa...

Page 16: ... 0 close dynamic filter 1 1 digital filter strength 2 2 digital filter strength 3 3 digital filter strength 4 4 digital filter strength 5 5 digital filter strength 6 6 digital filter strength Note Pls setting dynamic filter strength carefully the No is bigger more stable if the loaded weight shake not too much The setting is less than 3 C29 Noise filter option 0 close noise filter 1 1 digital filt...

Page 17: ...S W A S W B S W C X X X X X X X X X X X X C R C K S 1 2 3 4 5 6 State A Bits0 1 2 0 1 2 Decimal point position 1 0 0 XXXXXX0 0 1 0 XXXXXXX 1 1 0 XXXXX X 0 0 1 XXXX XX 1 0 1 XXX XXX Bits3 4 Division 0 1 X1 1 0 X2 State B BitsS function Bits0 gross 0 net 1 Bits1 symbol positive 0 negative 1 Bits2 overload or lower zero 1 Bits3 dynamic 1 ...

Page 18: ... unit lb 0 kg 1 Bits5 Constant 1 Bits6 Constant 0 State C Bit2 Bit1 Bit0 unit 0 0 0 Kg or lb 0 0 1 g 0 1 0 t Bit 3 printing 1 Bit 4 Extend display 1 Bit 5 Constant 1 Bit 6 Constant 0 5 2 Print output format Date 12 09 01 Time 16 08 34 Gross 2800kg ...

Page 19: ...ed weight 2 check load cell connection 3 checking load cell check input and output resistance to judge it is good or not Display nnnnnn 1 calibration is no good 2 cell single line is connect a wrong line 3 the cell is bad 1 check scale is resisted or not foot is kept level or not 2 check load cell connection 3 checking load cell check input and output resistance to judge it is good or not ERR1 dur...

Page 20: ...eight and scale is stable start calibration ERR5 EEPROM check error change PCB 6 2 Daily maintenance 1 In order to ensure indicator display clearly and prolong use life the indicator should not be placed directly on sunlight 2 Load cell and indicator should be well connected the system should have a good ground away from strong electric field magnetic field 3 Do not use indicator outside in rainy ...

Page 21: ...ue 1 C04 Max capacity 10000 C05 Empty scales calibration 0 C06 Capacity calibration 0 C07 restore the default parameters 0 C08 Warning tone 1 C09 Automatic power off 0 C10 Power saving mode 0 C11 Hold function 0 C12 Animal weighing mode 0 C13 Upper limit warning 000000 C14 Lower limit warning 000000 C15 Inner code display C16 Date C17 Time C18 Serial interface data output methord 0 C19 Serial inte...

Page 22: ...of 21 C22 Automatic zero tracking range 0 5 C23 Automatic zero tracking time 1 C24 Verload range 9 C25 Negative display range 10 C26 Standstill time 1 C27 Standstill range 2 C28 Dynamic filter 0 C29 Noisy filter 2 C30 C40 Reseverd menu ...