Operating Instructions Combics CAIXS2
67
Operation
Example
: 100% of a sample material should be weighed.
Configuration: The “Weighing in percent" application is selected, and printout has
been set up.
t
Place empty container on the scale.
)
.
t
Tare the scale.
This is not required if the automatic tare function is active. The tare weight is
saved automatically when you place the container on the platform.
t
Add reference material to the container in accordance with reference
percentage value (in this example, 85 g).
O
.
t
Start the calculation of the reference weight by pressing the
O
key.
y
The calculation is based on the active net weight value and the reference
percentage value entered.
If the weight is too light, an error code is shown in the main display
Inf 29
.
If this is the case, set the minimum load to a smaller number of digits.
t
Add additional material until the reference percentage value has been reached
(in this example, 100 g).
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...