Operating Instructions Combics CAIXS2
77
Operation
Classification
W
With this application, you can determine whether the weight of a given sample lies
within the limits of a defined weight class (
Applic.2
menu).
Features
– Classification with 3 or 5 weight classes.
Setting:
Applic./Applic.2/CLASS./PARAM.2/QTY.
menu item 4.8.
– Enter the upper limits of weight classes using the keypad or by saving weight
values from a load on the platform.
– Enter the upper limits of weight classes as absolute values or as a percentage of
deviation from the upper limit of Class 1.
Setting:
Applic./Applic.2/CLASS./PARAM.2/INPUT
menu item 4.9.
– Activate Info mode with the
I
key.
– Toggle the main display between classification display and weight display by
pressing the
w
key.
– Automatic results printout .
Setting:
Applic./Applic.2/CLASS./PARAM.2/PRINT
menu item 4.10.
– Automatic taring of container weight.
Setting:
Applic./aut.tare
, menu item 3.7.
– Automatic initialization when the scale is switched on.
Setting:
Applic./aut.strt
, menu item 3.8.
Exit Application, Delete Parameters
The initialization values remain active until they are deleted using the
c
key,
overwritten, or until the application is changed. The class limits also remain saved
after the scale is turned off.
You can assign different functions to the
c
to delete applications. When you
clear applications, you can delete either the data stored for all applications or just
selected data stored for the active application.
Setting:
APPLIC/CLER.CF
menu item 3.24.
Tare Function:
If you store a tare (weight value) by pressing the
)
key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting:
APPLIC./TARE.FNC
menu item 3.25.1 (factory setting).
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting:
Applic./TARE.FNC
menu item 3.25.2.
Restore factory default settings:
Applic./Reset
menu item 9.1.
Delimiters
To use the Classification application, you need to enter the delimiters that separate
one class from another. Limits between the individual weigh classes are required for
the classification. The lower limit of Class 1 is defined by the preset minimum load.
The other classes are configured by defining their upper limits. There are two ways
to enter the delimiters:
By
saving the weight value indicated
: Each upper limit value, with the exception
of the highest class, is entered using the keypad or by saving the weight value of a
load on the weighing platform.
Summary of Contents for Combics 2 CAIXS2
Page 120: ...120 Operating Instructions Combics CAIXS2 Declarations of Conformity...
Page 121: ...Operating Instructions Combics CAIXS2 121 Declaration of Conformity...
Page 123: ...Operating Instructions Combics CAIXS2 123...
Page 149: ...Operating Instructions Combics CAIXS2 149...
Page 172: ...172 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 173: ...Operating Instructions Combics CAIXS2 173 IECEx Certificate...
Page 174: ...174 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 175: ...Operating Instructions Combics CAIXS2 175 IECEx Certificate...
Page 176: ...176 Operating Instructions Combics CAIXS2 IECEx Certificate...
Page 178: ...178 Operating Instructions Combics CAIXS2...